A diagnostic front end with a noise floor problem
An early-stage medical diagnostics company
recovered to bench
Noise floor
2
Board revisions
40
Units in pilot build
The problem
A benchtop prototype produced clean readings. The same circuit, once integrated into the product enclosure alongside the power supply and display, lost roughly a decade of usable resolution — enough to make the device useless for its intended assay.
What we did
- 01
Reproduced the fault in a controlled setup to establish that the coupling was conducted through the supply rather than radiated.
- 02
Redesigned the board with a separated analogue section, a quiet reference and a deliberate single-point return, rather than adding shielding after the fact.
- 03
Moved the switching supply off the sensitive board entirely and specified a filtered feed.
- 04
Built a functional test that measures the noise floor on every unit, so a regression cannot silently reach a customer.
Outcome
The integrated device recovered its bench performance, and the per-unit noise floor test became part of the standard build. The design change was documented as it happened, which kept the validation burden contained.